Hach TL2100N laboratory turbidimeter, previously used and serviced under an annual maintenance contract with the manufacturer. It was last serviced and calibrated in 2018, then replaced in service with a newer model. The instrument powers on and appears to function, though recalibration will be required before use in lab-accurate or regulatory applications.
Key Features & Specifications:
- Measurement range: 0 to ~4,000 NTU with automatic range selection
- Accuracy: ±2% of reading or ±0.01 NTU (0–1,000 NTU range)
- Repeatability: ±1% of reading or ±0.01 NTU
- Light Source: Tungsten-filament lamp
- Sample Cell Compatibility: 25 mm round sample cells (adapters available)
- Interface / Output: RS-232 serial output
- Power: 115/230 VAC, 50/60 Hz
Condition & What’s NOT Included:
- Condition: Used. Last serviced/calibrated in 2018; turns on but requires recalibration.
- Missing: Does not include turbidity cells or calibration standards.
- Included: Power cable only.